Drift characteristics of the PE6000 ICPMS for solution analysis of geological samples
M. Norman, GEMOC, Macquarie
One of the principal limitations on the accuracy of an ICPMS determination
is the drift of the instrument response relative to the calibration
standard during the course of a run. This drift is often a complex
function of mass and can be the single most important factor in
determining the accuracy of an analysis (Eggins et al., 1997,
Chemical Geology 134, 311-326). This is especially a problem for
geological samples with their inherently complex matrix, and usually
requires addition of multiple internal standards to the analysed
solution to provide an adequate correction.
This abstract describes the drift characteristics observed for
two sets of whole rock analyses conducted recently (April 22-24,
1997) on the Perkin Elmer 6000 ICPMS at Macquarie University.
Samples consisted of 100 mg of basalt, pelagic sediments, and
ultramafic xenoliths dissolved in concentrated HF+HNO3. A mixed
internal standard consisting of 6Li, As, Rh, In, Re, and Bi was
added, and the solution brought to a final volume of 100 ml with
2% HNO3. The nominal dilution factor of the final solutions were
1:1000, and nominal concentrations of the internal standards were
30 ppb Li and 10 ppb for the other elements. Both runs were calibrated
against the well characterized basalt standard BHVO-1, and another
basalt from Kilauea volcano (KIL1489) was run peridocially as
a drift monitor.
Figures 1 and 2 show the intensities of each analysis of KIL1489
normalised to the first analysis of this sample in each run. These
data show that a significant loss of intensity for the light masses
occurred early in each run, which appears to be a fairly typical
pattern for solution ICPMS analysis regardless of the type of
instrument. In these experiment, the heavier masses show less
overall change in signal intensity, although the drift is clearly
a non-linear function of both mass and time
Drift corrections that simply normalize a group of elements to
a single internal standard are clearly inappropriate if better
than 10-20% accuracy is required, as it is in most modern geochemical
research problems. Fortunately, these data suggest that the response
characteristics for the PE 6000 are relatively smooth over limited
regions of the mass range, so that corrections based on interpolations
between the internal standard masses should improve both the accuracy
and the precision based on replicate determinations of a single
sample, although additional corrections based on a drift monitor
analysed every few samples may be necessary to achieve the highest
levels of accuracy and precision.
Understanding, controlling, and correcting for instrument drift
such as this should be a high priority of all manufacturers of
ICPMS instrumentation.
Back to the GEMOC Abstract Titles Page
© Copyright Macquarie University | Privacy Statement | Accessibility Information
Site Publisher: DVC Development and External Relations | Last Updated: 18 August 2008
ABN 90 952 801 237 | CRICOS Provider No 00002J